REFERENCE SAMPLE SO-1

REFERENCE SAMPLE SO-1
Standard samples SO-1 are designed to determine the conditional sensitivity, check the resolution and error of the depth gauge of an ultrasonic flaw detector.
Specifications:
1 – made of organic glass according to GOST 17622;
2 – speed of propagation of a longitudinal ultrasonic wave at a frequency of 2.5 ± 0.2 MHz at a temperature of 20 ± 5 ° C
is 2670±133m/s;
3 – propagation time of ultrasonic vibrations in the forward and reverse directions 20 ± 1 μs;
4 – geometric dimensions comply with the requirements of GOST 14782.
Contents of delivery:
Sample SO-1, verification certificate.